Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Tophit CPS491- Analog non-glass pH electrode for heavily soiled media

Analog non-glass pH sensor
Tophit CPS491

ISFET electrode for heavily polluted media in chemical processes, paper or paint production

CPS491
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
from CHF ??.-
  • Measuring principle

    ISFET

  • Application

    - process applications - non-glass, break-proof - low temperatures - min. conductivity >500µS/cm - emulsions, suspensions, organic solvents

  • Characteristic

    - rapid response at low temperatures - stabilised gel insensitive to contamination - lower amount of maintenance required than with glass - overhead installation

  • Measurement range

    pH 0-14

  • Measuring principle

    - open aperture - double gel reference

  • Design

    - all shaft lengths with temperature sensor - advanced gel technology

  • Material

    PEEK / metal oxide / Perfluorelastomere

  • Dimension

    Diameter: 12 mm (0.46 inch)
    Shaft lengths: 120, 225, 360 and 425 mm
    (4.68,8.77,14.04 and 16,57 inch)

  • Process temperature

    max. 110°C
    (230°F)

  • Process pressure

    max. 10 bar
    (145 psi)

  • Temperature sensor

    Optional with integrated Pt1000.

  • Ex certification

    ATEX
    FM

  • Connection

    Top68 connection head

  • Ingres protection

    IP68

Configure to download the CAD drawings of this product