Zoek op trefwoord, order- of productcode of serienummer, bv. “CM442” of “Technische informatie”
Voer tenminste 2 karakters in om het zoeken te starten.
  • Geschiedenis

Digital conductivity sensor
Memosens CLS15E

Memosens CLS15E - Digital contacting conductivity sensor for pure and ultrapure water ©Endress+Hauser
Memosens CLS15E - Digital conductivity sensor for pure water and ultrapure water applications ©Endress+Hauser

The conductivity of a liquid can be measured using the conductive or toroidal measuring principles. This video shows what it is about and how these measuring principles work.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens 2.0 contacting conductivity sensor for standard applications in pure and ultrapure water

Memosens CLS15E is perfectly suited for conductivity measurement in applications with low measuring ranges such as boiler feedwater and chip cleaning. The contacting conductivity sensor performs reliably and accurately even in hazardous areas. CLS15E features Memosens 2.0 digital technology, offering extended storage of calibration, adjustment and process data. It facilitates predictive maintenance and provides the perfect basis for IIoT services.

  • Eenvoudige producten

  • Gemakkelijk te selecteren, installeren en gebruiken

Technische topkwaliteit

Eenvoud

  • Standaardproducten

  • Betrouwbaar, robuust en onderhoudsarm

Technische topkwaliteit

Eenvoud

  • High-end producten

  • Functioneel en uiterst doelmatig

Technische topkwaliteit

Eenvoud

  • Gespecialiseerde producten

  • Ontworpen voor veeleisende applicaties

Technische topkwaliteit

Eenvoud

Variabel

FLEX selectie Technische topkwaliteit Eenvoud
  • F
  • L
  • E
  • X

Fundamental selectie

Voldoe aan uw basismeetvoorwaarden

Technische topkwaliteit
Eenvoud
  • F
  • L
  • E
  • X

Lean selectie

Beheers eenvoudig uw belangrijkste processen

Technische topkwaliteit
Eenvoud
  • F
  • L
  • E
  • X

Extended selectie

Optimaliseer uw processen met innovatieve technologieën

Technische topkwaliteit
Eenvoud
  • F
  • L
  • E
  • X

Xpert selectie

Beheers uw meest uitdagende applicaties

Technische topkwaliteit
Eenvoud

Variabel

  • Overzicht specificaties

    • Measurement range

      k=0,01: 0.04 to 20 µS/cm
      k=0,1: 0.10 to 200 µS/cm

    • Process temperature

      Threaded with fixed cable:
      -20 to 100 °C (-4 to 212 °F)
      Threaded with plug-in head:
      -20 to 120 °C (-4 to 248 °F)
      Sterilization: max. 140 °C (284 °F) for 30 minutes

    • Process pressure

      13 bar at 20 °C (188 psi at 68 °F) absolute
      1 bar at 120 °C (14 psi at 248 °F) absolute

please wait Loading alternative products

Snelle selectie

Kies uit één van de voorgestelde opties
Verkoopprijs
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

of

Individuele configuratie

Ga verder met uw keuze uit alle mogelijke opties
  • Voordelen

    • Designed for low maintenance and a long operating life, the sensor offers best value for money.

    • The sensor is sterilizable and autoclavable, allowing for installation in sterile plants.

    • Thanks to its electrode geometry, Memosens CLS15E provides...

  • Toepassingsgebied

    Memosens CLS15E measures conductivity in pure and ultrapure water applications for the monitoring and control of:

    • Boiler feedwater

    • Ion exchangers

    • Reverse osmosis

    • Cooling water

    • Distillation

    • Chip cleaning

    The digital sensor can be used in power plants for:

    • Degassed conductivity

    • pH measurement with...

Kenmerken en specificaties

Terug

Details

  • Meetprincipe

    Conductive

  • Application

    Measurement in pure and ultrapure water range

    Monitoring of ion exchangers

    Reverse osmosis

    Distillation and chip cleaning

  • Characteristic

    Digital 2-electrode conductivity sensor

  • Measurement range

    k=0,01: 0.04 to 20 µS/cm
    k=0,1: 0.10 to 200 µS/cm

  • Measuring principle

    Conductive conductivity cell with electropolished stainless steel electrodes

  • Design

    2-electrode conductivity cell with coaxially arranged electrodes, electropolished

  • Material

    Electrode: stainless steel 1.4435
    Electrode shaft: PES

  • Dimension

    Electrode diameter: 16 mm (0.63 inch)
    Electrode length: appr. 55 mm (2.17 inch)

  • Process temperature

    Threaded with fixed cable:
    -20 to 100 °C (-4 to 212 °F)
    Threaded with plug-in head:
    -20 to 120 °C (-4 to 248 °F)
    Sterilization: max. 140 °C (284 °F) for 30 minutes

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F) absolute
    1 bar at 120 °C (14 psi at 248 °F) absolute

  • Temperature sensor

    Pt1000

  • Ex certification

    ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex

  • Connection

    Process: 1/2" and 3/4" NPT, Clamp 1 1/2"
    Sensor connection: Inductive, digital connection head with Memosens 2.0 technology

  • Ingres protection

    IP68

  • Additional certifications

    Calibration certification of the cell constante and temperature, material certification 3.1

Documenten / handleidingen / software

Accessoires / Spare parts

Nr
Order code
Specification
    • Order code 50004381

      PVDF adapter voor CLS15 en CLS19

      Adapted voor sensoren met proceskoppel- ing G1/2"-G1".

    • Order code 50004738

      AM32 PVC-ADAPTER

      For conductivity cell installation into T-90-piece or cross piece DN 32

    • Order code 50004739

      AM40 PVC-ADAPTER

      Voor geleidbaarheidscelinstallatie in T-90-stuk of kruisstuk DN40

    • Order code 50004740

      AM50 PVC-ADAPTER

      For conductivity cell installation into T-90-piece or cross piece DN50

    • Order code 50066536

      PVC LIJMSTUK

      Met G½" binnendraad. Om CLS15/19 in DN20 PVC T-stuk te lijmen

    • Order code 50081902

      Kalibratieoplossing CLY11-A, 74,0 uS/cm

      inhoud: 500 ml volgens DIN IEC746-3, ASTM1125-91_

    • Order code 50081903

      KALIBRATIEVLOEISTOF CLY11-B, 149,6 uS

      inhoud: 500 ml volgens DIN IEC746-3, ASTM1125-91

    • Order code 50081904

      Kalibratieoplossing CLY11-C,1,406mS/cm

      inhoud: 500 ml volgens DIN IEC746-3, ASTM1125-91

    • Order code 50081905

      Kalibratievloeistof CLY11-D, 12,64 mS/cm

      500 ml, referentietemperatuur 25°C. Precisie-oplossing bedoeld SRM naar NIST. Met temp. tabel voor gekwalificeerde ijking van de geleidbaarheid meetsystemen volgens ISO

    • Order code 50081906

      Kalibreervloeistof CLY11-E,107,00 mS /cm

      inhoud: 500 ml volgens DIN IEC746-3, ASTM1125-91