Specs at a glance
Measurement range
Trace: 0 to 5 mg/l ClO2 Standard: 0 to 20 mg/l ClO2 High: 0 to 200 mg/l ClO2
Process temperature
0 to 55 °C , non-freezing (32 to 130 °F)
Process pressure
Max. 2 bar abs (Max. 29 psi abs)
Field of application
Memosens CCS50D is a robust, low maintenance sensor for chlorine dioxide measurement. It provides stable and fast measured values in drinking water, process water and utilities. The sensor can be used to ensure efficient disinfection for highest water quality and safety, to avoid overdosing of chlorine dioxide or to ensure the absence of ClO2 in beverage plants and reverse osmosis. Memosens CCS50D is not the latest Memosens generation. More information on the new Memosens CCS50E sensor here .
Memosens CCS50D is a chlorine dioxide sensor for skid builders and end customers. It measures reliably in:
Drinking water to ensure sufficient disinfection
Cooling systems or towers to avoid pathogen growth
Process water - for hygienic and safe packaging and bottling of products
Beverage plants to secure the absence of chlorine dioxide
Benefits
The right sensor version for every application: From trace measurement up to chlorine dioxide concentrations of 200 mg/l.
Fast response time provides accurate process view and enables prompt reaction to process changes as well as efficient process control.
Increased process safety: precise and long-term stable measurement ensures consistent process monitoring and allows for lowest disinfectant concentration.
Flexible installation: sensor can be installed in the CCA151 and CCA250 flow assemblies or in immersion assemblies. Measurement virtually flow independent at flow velocities above 5 l/h (CCA151), 30 l/h (CCA250) or above 15 cm/s (immersion).
More process up-time thanks to fast sensor exchange: precalibrate the sensor in your lab and then swap it into your process with plug & play.
Connection to the Liquiline multiparameter transmitter allows easy combination with other relevant parameters of liquid analysis.
Extended selection
High-end products
Highly functional and convenient
Technical excellence
Simplicity
©Endress+Hauser
Simple products
Easy to select, install and operate
Technical excellence
Simplicity
Standard products
Reliable, robust and low-maintenance
Technical excellence
Simplicity
High-end products
Highly functional and convenient
Technical excellence
Simplicity
Specialized products
Designed for demanding applications
Technical excellence
Simplicity
FLEX selections
Technical excellence
Simplicity
Fundamental selection
Meet your basic measurement needs
Technical excellence
Simplicity
Lean selection
Handle your core processes easily
Technical excellence
Simplicity
Extended selection
Optimize your processes with innovative technologies
Technical excellence
Simplicity
Xpert selection
Master your most challenging applications
Technical excellence
Simplicity
Loading alternative products
Success stories
An automated system for pH measurements at AVR
AVR Waste in Botlek, Rotterdam, processes up to 1.7 million tons of residual waste per year. Discover how they optimized their pH measurement thanks to Endress+Hauser
Read the full story
Pure high quality beer and clean wastewater
Discover how Heineken meets all regulations to protect the brazilian environment with the Viomax CAS51D digital nitrate or SAC sensor from Endress+Hauser.
Read the full story
Smart vertical farming in big cities
Intuitive pH and conductivity measurement in smart vertical farming plants to meet quality requirements and ensure comprehensive
monitoring of processes.
Read the full story
Application
Adequate disinfection in drinking water, preventing of pathogen formation in cooling water systems, water used to wash pre-packaged vegetables, ensures the absence of chlorine dioxide in beverage systems
Characteristic
Amperometric measurement of dissolved chlorine dioxide
Measurement range
Trace: 0 to 5 mg/l ClO2 Standard: 0 to 20 mg/l ClO2 High: 0 to 200 mg/l ClO2
Measuring principle
Closed, membrane covered measuring cell Reduction of chlordioxide (ClO2) to chloride at the cathode with appr. 120mV
Design
Closed amperometric 2-electrode measuring cell with PVDF membrane
Material
Sensor shaft: POM Membrane: PVDF Membrane cap: PVDF
Dimension
Diameter: 25 mm (0.98 inch) Length: 161 mm (6.34 inch)
Process temperature
0 to 55 °C , non-freezing (32 to 130 °F)
Process pressure
Max. 2 bar abs (Max. 29 psi abs)
Temperature sensor
10k NTC integrated (Memosens)
Connection
Inductive, digital connection head with Memosens
Most downloaded
CCS50D
English version - 07/2020
Digital sensor with Memosens technology for determining chlorine dioxide
EN
BG CS DA DE EL ES ET FI FR HR HU ID IT JA KO LT LV NL NO PL PT RO RU SK SL SV TH TR ZH
20/07/2020
08/02/2019
19/05/2017
Language:
English
Version:
20/07/2020
File size:
1.4 MB
File name:
BA01773CEN_0320.pdf
118 documents
Memosens CCS50D
Technical Information
Digital sensor with Memosens technology for determining chlorine dioxide
EN
German Spanish French Italian Japanese Polish Portuguese Russian
08/02/2019
19/05/2017
Language:
English
Version:
08/02/2019
File size:
724 KB
File name:
TI01353CEN_0219.pdf
Digital sensor with Memosens technology for determining chlorine dioxide
EN
Bulgarian Czech Danish German Greek Spanish Estonian Finnish French Croatian Hungarian Indonesian Italian Japanese Korean Lithuanian Latvian Dutch Norwegian Polish Portuguese Romanian Russian Slovak Slovenian Swedish Thai Turkish Chinese
20/07/2020
08/02/2019
19/05/2017
Language:
English
Version:
20/07/2020
File size:
1.4 MB
File name:
BA01773CEN_0320.pdf
Disinfection applications
Measuring ranges and suitable sensors
Language:
English
Version:
16/06/2023
File size:
603.1 KB
File name:
PU01405CEN-disinfection-sensor-application.pdf
Messbereiche und geeignete Sensoren
Language:
German
Version:
03/11/2023
File size:
678.7 KB
File name:
PU01405CDE_02.23_Desinfektionsanwendungen.pdf
Kompetenz in der Flüssigkeitsanalyse
Vom Sensor bis zur schlüsselfertigen Komplettlösung
Language:
German
Version:
21/09/2022
File size:
1.3 MB
File name:
FA01018CDE_0422_Analyse_Portfolio_lowres.pdf
Sensors, transmitters, compact devices and assemblies
Experts in Liquid Analysis
Language:
English
Version:
08/03/2022
File size:
4.8 MB
File name:
FA00007CEN_Analyse_20.22_LowRes.pdf
Sensoren, Messumformer, Kompaktgeräte and Armaturen
Kompetenz in der Flüssigkeitsanalyse
Language:
German
Version:
08/03/2022
File size:
4.6 MB
File name:
FA00007CDE_Analyse_20.22_LowRes.pdf
Aus Freude an gutem Wasser
Frische Ideen für Wasserwerke, Verteilernetze, Oberflächenwasser und Industrieanlagen
Language:
German
Version:
14/07/2023
File size:
2.1 MB
File name:
CP01335CDE_0223_Kompetenz in Wasser.pdf
For the love of quality water
Fresh ideas for waterworks, distribution networks, surface waters and industrial plants.
Language:
English
Version:
14/07/2023
File size:
2 MB
File name:
CP01335CEN_0223_Competence in water_LoRes.pdf
Chlordioxid-Analyse in Trink- und Prozesswasser
Sichere, effektive Desinfektion mit Memosens CCS50D
Language:
German
Version:
27/10/2017
File size:
743.8 KB
File name:
IN01114CDE_0117.pdf
Chlorine dioxide measurement in drinking and process water
Safe and efficient disinfection with Memosens CCS50D
Language:
English
Version:
27/10/2017
File size:
644.7 KB
File name:
IN01114CEN_0117.pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Turkish
Version:
17/10/2023
File size:
185.9 KB
File name:
MSDS_CCV05-50x_51x_(TR).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
German
Version:
19/02/2024
File size:
42.3 KB
File name:
MSDS_CCV05-50x_51x_(D).pdf
CCV05-50x_51x (Czech Republic)
Electrolyte for CCS50(D)/CCS51(D)
Language:
Czech
Version:
29/04/2022
File size:
222.3 KB
File name:
MSDS_CCV05-50x_51x_(CZ).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
French
Version:
29/04/2022
File size:
42.9 KB
File name:
MSDS_CCV05-50x_51x_(F).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Spanish
Version:
29/04/2022
File size:
42.4 KB
File name:
MSDS_CCV05-50x_51x_(E).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Estonian
Version:
29/04/2022
File size:
40.4 KB
File name:
MSDS_CCV05-50x_51x_(EST).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Greek
Version:
29/04/2022
File size:
199.4 KB
File name:
MSDS_CCV05-50x_51x_(GR).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
40.1 KB
File name:
MSDS_CCV05-50x_51x_(M).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
40.3 KB
File name:
MSDS_CCV05-50x_51x_(IRL).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
German
Version:
29/04/2022
File size:
42.7 KB
File name:
MSDS_CCV05-50x_51x_(A).pdf
CCV05-50x_51x (Belgium (Dutch))
Electrolyte for CCS50(D)/CCS51(D)
Language:
Dutch
Version:
29/04/2022
File size:
42.1 KB
File name:
MSDS_CCV05-50x_51x_(BNL).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
33.9 KB
File name:
MSDS_CCV05-50x_51x_(CDN).pdf
CCV05-50x_51x (Switzerland)
Electrolyte for CCS50(D)/CCS51(D)
Language:
German
Version:
29/04/2022
File size:
42.5 KB
File name:
MSDS_CCV05-50x_51x_(CH).pdf
CCV05-50x_51x (Saudi Arabien)
Electrolyte for CCS50(D)/CCS51(D)
Language:
Arabic
Version:
29/04/2022
File size:
168.8 KB
File name:
MSDS_CCV05-50x_51x_(SAA).pdf
CCV05-50x_51x (European Union)
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
40.1 KB
File name:
MSDS_CCV05-50x_51x_(EU).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Japanese
Version:
29/04/2022
File size:
273.5 KB
File name:
MSDS_CCV05-50x_51x_(J).pdf
CCV05-50x_51x (New Zealand)
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
37.8 KB
File name:
MSDS_CCV05-50x_51x_(NZ).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Spanish
Version:
29/04/2022
File size:
38.7 KB
File name:
MSDS_CCV05-50x_51x_(MEX).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Chinese
Version:
29/04/2022
File size:
257.6 KB
File name:
MSDS_CCV05-50x_51x_(RC).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Polish
Version:
29/04/2022
File size:
188.1 KB
File name:
MSDS_CCV05-50x_51x_(PL).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Slovak
Version:
29/04/2022
File size:
220.6 KB
File name:
MSDS_CCV05-50x_51x_(SK).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Russian
Version:
29/04/2022
File size:
187.3 KB
File name:
MSDS_CCV05-50x_51x_(RUS).pdf
CCV05-50x_51x (United Arab Emirates)
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
36.5 KB
File name:
MSDS_CCV05-50x_51x_(UAE).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Unknown
Version:
29/04/2022
File size:
180.1 KB
File name:
MSDS_CCV05-50x_51x_(UK).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Danish
Version:
29/04/2022
File size:
41.3 KB
File name:
MSDS_CCV05-50x_51x_(DK).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Croatian
Version:
29/04/2022
File size:
199.7 KB
File name:
MSDS_CCV05-50x_51x_(CRO).pdf
CCV05-50x_51x (United Kingdom)
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
37.2 KB
File name:
MSDS_CCV05-50x_51x_(GB).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
Italian
Version:
29/04/2022
File size:
42 KB
File name:
MSDS_CCV05-50x_51x_(I).pdf
Electrolyte for CCS50(D)/CCS51(D)
Language:
English
Version:
29/04/2022
File size:
37.2 KB
File name:
MSDS_CCV05-50x_51x_(IND).pdf
Search spare parts by number in drawing
60 spare parts & accessories
60 spare parts & accessories
60 spare parts & accessories
Drawing number
Order code
2
CCV05-
Kit adapter CCS5x(D/E) for CCA250
Kit adapter CCS5x(D/E) for CYA112
Kit adapter CCS5x(D/E) for CYA27+CCA151
Kit adapter CCS5x(D/E) for CYA27-xxxB
photometer M+N PF-3, disinfection
We value your privacy
We use cookies to enhance your browsing experience, collect statistics to optimize site functionality, and deliver tailored advertisements or content.
By selecting "Accept all", you consent to our use of cookies.
For further details please review our cookie policy .
Customize
Accept only essential
Accept all