Liquiline System CA80FE - uiterst nauwkeurige procesanalyzer voor aluminiummetingen

IJzeranalyzer
Liquiline System CA80FE

Colorimetrisch systeem voor de bewaking van drinkwater, mineraalwater en proceswater

CA80FE
Please Wait (spinning wheel)
Prijzen worden geladen
Prijzen tijdelijk niet beschikbaar
Price on request
Bezig met activeren van prijs
Prijs: Informatie verstrekken
from CHF ??.-
Filter meting-/ toepassingstaak
Meet-/toepassingstaak

Configureer om de CAD-tekeningen van dit product te downloaden

  • Meetprincipe

    Colorimetric

  • Characteristic

    Process analyzer for iron in aqueous solutions
    Comply with standard colorimetric measuring principle, ferrozine method

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39.2 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)
    Outdoor version: - 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar

  • Sample flow rate

    Min. 5 ml/min (0.17 fl.oz/min)

  • Consistency of the sample

    Suspended solids content
    Turbitity < 50 NTU, aqueous, homogenized

  • Application

    Monitoring of the iron content of potable water, mineral water and source water for industrial processes

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1 or 2 measuring channel
    1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0.05 to 2 mg/l Fe
    0.1 to 5 mg/l Fe
    0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe

  • Consumables

    Reagents and standard solutions CY80FE for the operation
    Regular maintenance is done with the parts of the maintenance kit CAV800